    摘要:

    为建立准确、高效的1,3,5-三氯-2,4,6-三硝基苯(TCTNB)纯度质量控制标准,系统对比了经典酸碱滴定法与新建立的高效液相色谱法(HPLC)的综合性能。酸碱滴定法通过氢氧化钠水解TCTNB后返滴定过量碱进行定量;HPLC采用低毒性甲醇替代传统有毒溶剂,通过C18色谱柱实现了目标物与杂质的基线分离。实验结果表明,酸碱滴定法操作繁琐(单次检测耗时5.9 h),且准确度严重不足,其测定均值(93.89%)较样品标称值存在高达-6.05%的系统性负偏差。相比之下,HPLC分析效率更高(单次耗时3.2 h),精密度良好(相对标准偏差<0.30%,n=6),测定均值(99.82%)与标称值(99.94%)高度吻合,展现出卓越的准确性。滴定法因水解后溶液颜色深导致终点判断主观误差大,且水解反应不完全或存在干扰性副反应是造成系统性负偏差的根本原因。HPLC凭借其优异的分离选择性和检测稳定性,有效克服了上述缺陷,在准确度、精密度、效率及绿色安全方面均显著优于传统滴定法。

    Abstract:

    To establish accurate and efficient quality control standards for the purity of 1,3,5-trichloro-2,4,6-trinitrobenzene (TCTNB), the comprehensive performance of the classical acid-base titration method and a newly developed green liquid chromatography method was systematically compared. The acid-base titration method quantifies TCTNB by hydrolyzing it with sodium hydroxide followed by back-titration of the excess base. The liquid chromatography method innovatively employed low-toxicity methanol instead of traditional toxic solvents and achieved baseline separation of the target compound from impurities on a C18 column. Experimental results indicated that the titration method was cumbersome (requiring 5.9 hours per analysis) and exhibited significant inaccuracy, showing a mean measured value of 93.89% with a substantial systematic negative bias of -6.05% compared to the sample''s nominal value. In contrast, the liquid chromatography method demonstrated higher efficiency (requiring 3.2 hours per analysis), good precision (with a relative standard deviation <0.30%, n=6), and excellent accuracy, showing a mean value of 99.82% that closely matched the nominal value of 99.94%. The substantial systematic error in the titration method was primarily attributed to the deep color of the hydrolyzed solution, which introduced subjectivity in endpoint determination, and potentially incomplete hydrolysis or interfering side reactions. The liquid chromatography method effectively overcame these limitations due to its superior separation selectivity and detection stability. The newly established liquid chromatography method significantly outperforms the traditional titration method in terms of accuracy, precision, efficiency, and green safety.

    参考文献
    相似文献
    引证文献
文章指标
  • PDF下载次数:
  • HTML阅读次数:
  • 摘要点击次数:
  • 引用次数:
引用本文

陈玲,冯志强,李怀姣,等.液相色谱法与酸碱滴定法在TCTNB纯度分析中的对比[J].含能材料, 2025, 33(10):1192-1200. DOI:10.11943/CJEM2025170.
CHEN Ling, FENG Zhi-qiang, LI Huai-jiao, et al. Comparative Application of Liquid Chromatography and Acid-Base Titration for TCTNB Purity Analysis[J]. Chinese Journal of Energetic Materials, 2025, 33(10):1192-1200. DOI:10.11943/CJEM2025170.

复制
历史
  • 收稿日期: 2025-07-26
  • 最后修改日期: 2025-10-28
  • 录用日期: 2025-10-22
  • 在线发布日期: 2025-10-22
  • 出版日期: